Intrinsic detector sensitivity analysis as a tool to characterize ArcC by Thahabah Alharthi, Phil Vial et al

Abstract Purpose: To investigate intrinsic sensitivity of an electronic portal imaging device (EPID) and the ArcCHECK detector and to use this in assessing their performance in detecting delivery variations for lung SBRT VMAT. The effect of detector spatial resolution and dose matrix interpolation on the gamma pass rate was also considered. Materials and methods: Fifteen patients’ lung SBRT VMAT plans were used. Delivery variations (errors) were introduced by modifying collimator angles, multi-leaf collimator (MLC) field sizes and MLC field shifts by ±5, ±2, and ±1 degrees or mm (investigating 103 plans in total). EPID and ArcCHECK measured signals with introduced variations were compared to measured signals without variations (baseline), using OmniPro-I mRT software and gamma criteria of 3%/3 mm, 2%/2 mm, 2%/1 mm, and 1%/1 mm, to test each system s basic performance. The measurement sampling resolution for each was also changed to 1 mm and results compared to those with the de