comparemela.com

Latest Breaking News On - Dose volume histogram - Page 1 : comparemela.com

Remote EPID-based dosimetric auditing using DVH patient dose analysis by P B Greer, T Standen et al

Objective. The aim of this work was to develop and validate a method for remote dosimetric auditing that enables dose-volume histogram parameter comparisons of measured and planned dose in the patient CT volume. Approach. The method is derived by adapting and combining a remote electronic portal imaging (EPID) based auditing method (Virtual Epid based Standard Phantom Audit VESPA) and a method to estimate 3D in-patient dose distributions from planar dosimetric measurements. The method was tested with a series of error-induced plans including monitor unit and multileaf collimator (MLC) positioning errors. A pilot audit study was conducted with eleven radiotherapy centres. IMRT plans from two clinical trials, a post-prostatectomy (RAVES trial) plan and a head and neck (HPV trial) plan were utilized. Clinically relevant DVH parameters for the planned dose and estimated measured dose were compared. Main results. The method was found to reproduce the induced dose errors within 0.5% and was